-   Home Page Field Missions Engineering More Info [MAP Menu]

molecule.list

Specify bulb contents

Syntax:


manifold bulb size molecule balance balance-chi [ tracer tracer-chi ]

Options:

The following parameters should be specified on each line:

manifold
The manifold to which the bulb is attached, integer number
bulb
The number of the bulb, integer number
size
The volume of the bulb, in liters
molecule
The molecule of interest which the bulb contains (standardized text field)
balance
The balance gas in the bulb
balance-chi
The ratio of molecule to balance gas
tracer
The tracer molecule (usually F12)
tracer-chi
The ratio of the tracer to balance gas (if present)

Examples:


man1  bulb2  4.5  ethane  N2  0.01  F12  0.005

Description:

molecule.list is used to specify the contents of all bulbs attached to the system for injection of reagents. This information is read by the drext and injected into the ASCII output stream, which is further read by dataRecv and then stored in "run.def" on the Linux side, which can then be used by the data analysis scripts.

Manifold and bulb numbers are standardized by the experimenter.

See also:



last updated: Thu Aug 7 14:42:53 2003 webmaster@huarp.harvard.edu
Copyright 2002 by the President and Fellows of Harvard College
[Home] [People] [More Info] [Research Areas] [Field Missions] [Engineering]